Skip to main content
× Home Services Products Catalogs Downloads TechSupport About Contact Careers (505) 872-0037

(505) 872-0037

Shopping Cart Icon
Cart
 
Login Icon
Login
Language Selector
en-fr
×
Americas
Europe
Middle East & Africa
Asia Pacific & Japan
Pfeiffer ASM340 Leak Detectors On Sale Ideal Vacuum XGC-320 Portable Digital Thermocouple Controllers On Sale Agilent IDP-15 Dry Scroll Vacuum Pumps On Sale Agilent TriScroll Pumps On Sale Edwards nXR Multi Stage Roots Pumps On Sale Agilent IDP-10 Dry Scroll Vacuum Pumps On Sale

Keyword       Part Number:      

× Vacuum Pumps Modular Vacuum Chambers Stainless Steel Vacuum Chambers Aluminum Vacuum Chambers ExploraVAC Unlimited Chambers ExploraVAC Vacuum Chambers Fittings and Flanges Feedthroughs Vacuum Valves Rebuild Kits, Parts, And Motors Vacuum Fluids, Oils, and Greases Turbo Pumps And Controllers Filters Traps and Silencers Convection and Vacuum Ovens Leak Detection and RGA Vacuum Pressure Measurement Recirculating Chillers And Water Baths
×

Load

Product Image 1
Product Image 2
Product Image 3

NEW Gasket Assembly Replacement for Agilent Variable Leak Valve, PN 9535050

Condition:
  New
Part Number:
  P105241
Warranty:
  Varian Warranty

Available Now   1  

Sale: €563.96

NEW Gasket Assembly Replacement for Agilent Variable Leak Valve, PN 9535050 563.96
Currency: Euro (Euro)

Description

Gasket Assembly Replacement Part for Agilent Variable Leak Valve PN: 9535050


This is the NEW Gasket Assembly Replacement Part for Agilent Variable Leak Valve (part number 951-5106). Please follow the instructions below to inspect, clean or replace the gasket assembly.

    How to Inspect, Clean or Replace gasket assembly:

  • Inspect the copper alloy gasket to be sure that it is clean, smooth, free of oxide, and protrudes above the surface of the gasket collar by at least 0.002 inch. A hand lens or microscope will facilitate inspection of the surface.
  • If the top surface of the gasket is scored, scratched, or nicked, the gasket assembly must be replaced.
  • If the copper portion of the assembly does not protrude at least 0.002 inch above the collar, the gasket assembly must be replaced.
  • Oily films or other residue should be removed using acetone.
  • Oxide or other slight surface imperfections can generally be removed by polishing with a very fine rouge paper. When polishing, rotate the gasket assembly about its center axis to avoid leaving scratches that cross the sealing surface. Clean the gasket with acetone after polishing.
  • Keep the gasket free of contamination while awaiting re-assembly.

We offer all the necessary tools to make repairs and adjustments to the variable leak valve below in our Related Products section, including replacement parts such as the complete repair/tool kit and gasket assemblies. Gasket Assembly Replacement Part, PN: 9535050

Brochure

Marca de agua con el logotipo de Ideal Vacuum
CONTACT US
Ideal Vacuum Products , LLC
5910 Midway Park Blvd NE
Albuquerque, New Mexico 87109-5805 USA

Phone: (505) 872-0037
Fax: (505) 872-9001
info@idealvac.com



Press ESC to Close


deepbox image 2
load time = 1.4121