× Bombas de vácuo Câmaras de vácuo modulares Câmaras de vácuo em aço inoxidável Câmaras de vácuo de alumínio ExploraVAC Câmaras Ilimitadas ExploraVAC Câmaras de Vácuo Conexões e Flanges Passatempos Válvulas de Vácuo Kits de reconstrução, peças e motores Vácuo Fluidos, Óleos e Graxas Turbobombas e controladores Filtros armadilhas e silenciadores Fornos de Convecção e Vácuo Detecção de vazamento e RGA Vácuo Pressão Medição Resfriadores e Banhos de Água Recirculantes
×

Load

Product Image 1
Product Image 2
Product Image 3

Substituição da tampa fria padrão Agilent Varian HS-32, peças da bomba de vácuo de difusão, 77252801

Doença:
  Novo
Número da peça:
  P107061
Garantia:
  Varian Warranty

Fora de estoque   

Oferta: €2,148.30

Substituição da tampa fria padrão Agilent Varian HS-32, peças da bomba de vácuo de difusão, 77252801 2148.3
Moeda: Euro (Euro)

Descrição

Substituição da bomba de difusão Agilent Varian HS-32 padrão Cold Cap.
Agilent Varian PN 77252801.

Essas tampas frias padrão da bomba de difusão Agilent Varian HS-32 são uma peça de reposição que pode ser instalada no local do cliente. Existem dois tipos de cold caps, padrão e Halo Baffle. O tipo de cold cap utilizado influenciará fatores como refluxo de óleo e velocidade final de bombeamento da bomba de difusão. Por exemplo, o Halo Baffle Cold Cap é uma opção que se encaixa dentro das bombas HS-32 no lugar do cold cap padrão. O Halo Baffle interrompe o backstream de forma tão eficaz quanto um defletor opticamente denso, mas retém 60% da velocidade da bomba. O backstreaming residual é tão baixo que não pode ser medido pelo método de coleta padrão da American Vacuum Society. Para a localização da tampa fria, consulte o item nº 1 na vista explodida do diagrama e na ilustração das peças mostrada abaixo. Para obter os procedimentos de remoção e instalação da tampa fria, consulte a página 38 do manual de instruções PDF abaixo. Essas tampas frias padrão são peças de reposição originais OEM Agilent Varian com número de peça 77252801.

Folheto

ITEM DESCRIPTION PART NO.
1 Standard Cold Cap Assembly 77252801
Water-cooled Halo Baffle K1856332
2
2a
2b
2c
Cold Cap Connectors & Gaskets
Gasket/Grommet (Kit of 4)L8839301
Nut & Follower (Brass)L8840301
Brass Female Coupling622445026
3 Jet Assembly 76511301
4 1 Buna O-ring for ASA inlet flange 78536002
5 Foreline Baffle Assembly
6 Centering ring for ISO foreline flange, 200K IC200SV
7 Sight Glass Service Kit for Pumps Built Before Oct. 1995:
3 Glass TubesF1755008
2 Brass Union 1/4 OD x 1/4 MPT622568108
2 Brass Elbow 69FL 6x4, 169VL6-4623451012
8 O-ring, Viton, Parker No. 2-011660892011
8 O-ring, Viton, Parker No. 2-111660892111
3 O-ring, Viton, Parker No. 2-221660892221
3 O-ring, Viton, Parker No. 2-213660892213
F6097301
7 Sight Glass Service Kit for Pumps Built After 1995 L9223001
8 Threaded Plugs, fill and drain 77261001
Hose Barb 1/2in Hose to 3/8in Pipe 624301111
9 Upper (water) Thermal Switch (Set at 220°F/104°C) K9050003
9 Lower (boiler) Thermal Switch (Set at 550°F/288°C) K9050004
10 Thermal Switch Brass Block F6253001
11 Heater wire, 10 AWG, 196 inches (2 per heater required) X3900-68003
11b Heater Wiring Connection Box
12 Heaters (six heaters required per pump, same voltage)
240V/415V/480V, 4000 watt Heater Element w/ leads647332010
200V, 4000 watt Heater Element647332020
Heater Retrofit for pumps built bef. May 1990
240V/451V/480V, 4000W Heater no leads (1 consists of 2-120V heater)647332075
Heater Kits (clamping plates not included)
240V Heater Retrofit Kit HS-32 PumpL6517308
415V Retrofit Kit HS-32 PumpL6517326
480V Retrofit Kit HS-32 PumpL6517309
13 Heater Crush Plate (Six required per pump) K7246001
14 Heater Clampling Plate (Six required per pump) K7247001
15 Heater Insulator (One required per pump) 75792001
16 Cover Plate 75791001
Heat Shield (Reusable) L6370001
O-Ring Kit for ASA Pumps includes:
1 Butyl O-ring for ASA inlet flange45390001
1 Butly O-ring foreline flange (2-2443)660893267
8 Viton O-rings fill and drain (2-213)660892213
4 Viton O-rings fill and drain (2-221)660892221
1 Sight glass O-ring (2-240)660892240
1 Graphite sight glass gasketL9228001
K0377167
DC-702 diffusion pump fluid, one-gallon 695472008
DC-704 diffusion pump fluid, one-gallon 695474008
DC-705 diffusion pump fluid, one-gallon 695475008
Installation and Operation Manual-download from www.agilent.com/chem/vacuum 699901140